The difference between Xenon Accelerated Weathring Chamber and UV Accelerated Weathering Tester
The difference between UV Chamber and Xenon Chamber
Coatings, plastics and other organic materials are exposed under natural climatic conditions and light
radiation over a period of time there will be loss of light, faded, yellowed, peeling, cracking, loss of
tensile strength and the whole layer off and so on. Even indoor light or also as a pigment or dye to
such material damage through the glazing sun. Thus for paint for outdoor use, such as building
exterior coatings and automotive coatings, weather resistance and light resistance are the most
important test items. Now there are many paint weathering resistance and light resistance test
method. While natural weathering test, a xenon arc lamp, a carbon arc lamp or light, ultraviolet light
irradiation, etc. artificial accelerated weathering tests commonly used method . While the most
common test equipment it’s UV Accelerated Weathering Tester and Xenon Weathering Accelerated
Aging Chamber, while these two equipment both simulated the Sunlight and have similar function but
it did different from each other. Today we are going to discuss the difference between two equipment
and how to choose right one according to your test standard.
UV mainly used to simulate the solar light in the ultraviolet part of the spectrum , by simulate the
ultraviolet radiation and condensation in the natural sunlight to carry out the accelerated weathering
tests on the materials. Fluorescent UV lamps similar to the common cold fluorescent lighting
electrically principle, but it can generate more ultraviolet light rather than visible or infrared light.
For different exposure applications, there are different types with different spectra of light available.
UVA340type
lamps to simulate sunlight well in the major shortwave length
ultraviolet spectral range.
UVA340
lamp spectral power distribution (SPD) and the solar spectrum from 360nm to dispose of
the spectrum is very similar. Artificial climate accelerated UVB
lamps are also commonly used lamp
aging test. Than UVA
lamps for material damage to speed faster, but shorter than 360nm
wavelength energy output of many materials can cause the actual results deviate.
Irradiance (light intensity) control to obtain accurate and reproducible results there is necessary. Most
UV aging test apparatus is equipped with irradiance control system. These precise irradiance control
system allows the user to select the measurement in a test irradiation. Feedback control system, the
irradiance can be continuously and automatically monitored and precisely controlled. Control system
by adjusting the lamp power is automatically on the illumination lamp due to aging or other causes of
inadequate compensation.
Fluorescent UV lamps due to their own inherent spectral stability make irradiance control simplified.
All light sources are aging with time becoming weaker. However, other types of fluorescent lamps is
different from its spectral energy distribution does not change with time. This feature improves the
reproducibility of the test results, which is a big advantage.
A trial showed that using a 2h lights and a lamp in use 5600h aging test system equipped with
irradiance control of the output power of no significant difference between the control device
irradiance light intensity can be maintained constant. In addition, the spectral distribution of the
energy they also show no change with a xenon arc lamp that is very different.
Xenon arc mainly through all analog spectrum (full spectrum) to reproduce the existence of different
environments destructive waves, can provide the appropriate environment for simulation and
accelerated testing of scientific research, product development and quality control. These two methods
are based on entirely different principles. Xenon lamp irradiation chamber reproduced in
whole solar spectrum, including ultraviolet, visible and infrared light, which is the object of the
simulated sunlight. The UV aging test does not attempt to imitate the sun's rays, but just imitate
destructive effect of sunlight. It is based on longterm
outdoor durability of materials exposed by shortwave
ultraviolet radiation damage caused by the aging of the maximum principle.
Two easy ways to different two equipment:
1.On function, Xenon Weathering Accelerated Aging Chamber can realize the function of adjusted
and controlled of temperature and humidity, light, rain test however UV Accelerated Weathering
Tester can only detect humidity and not able to adjust or control.
2. Irradiation intensity, Xenon Weathering Accelerated Aging Chamber it’s 0.46W/m21200W/
m2 while UV Accelerated Weathering Tester it’s 0.46W/m 21.2 W/m2.